Magnetic Resonance Imaging (MRI) Machine
791102,PASIGHAT
- Delivery
- 120 days
Specifications & requirements · 14
- Specification Name
- बड के िलए आवOयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Helium refill time should not be less than 2 years
- Yes,
- Magnet Strength
- 1.5 T
- Comprehensive Maintenance ( Minimum Percentage )
- 3 %
- Comprehensive Maintenance ( Maximum Percentage )
- 25 %
- Warranty of required product
- 5 Year
- Comprehensive Maintenance Duration (Post Warranty)
- 5 Year
- Magnet
- It should have at least 60 cm patient bore with flared opening. Magnet length should be 170 cm or less. Guaranteed magnet homogeneity shall be less than or equal to 0.5 ppm at 40 cm DSV and less than or equal to 5 ppm at 50 cm DSV.
- RF SYSTEM
- A fully digital RF system capable of transmitting power of at least 10 kw. It should have the technology that supports at least 32 channel with 32 ADC signal processing.
- GRADIENT SYSTEM
- The gradient should be activity shielded with a slew rate at least 125 T/m/s and a peak amplitude of at least 35 mT/m. Both Max slew rate and peak amplitude should be achieved simultaneously. Mention the rise time.
- COIL SYSTEM
- Head / Neck coil having at least 16 elements or more, capable of high- resolution neuro -vascular imaging. Spine coil having at least 24 elements or more for thoracic and lumbar spine imaging. Body coil with at least 30 channels in combination with spine coil capable of doing abdomen, pelvis, MRCP and Peripheral imaging.
- of Product Cost Payable on Product Delivery
- 70%
- Min Cost Allocation for ICT as a % of product cost
- 30%
- Number of days allowed for ICT after site readiness communication to seller
- 20 Days