Automated Blood Culture System
160016,GMSH-SECTOR 16, CHANDIGARH
- Delivery
- 60 days
Specifications & requirements · 41
- Specification Name
- बड के िलए आव8यक अनुमत मूHय /Bid Requirement (Allowed Values)
- Product Description
- Automated Blood Culture System
- Technology
- Fluorometric technology / Colorimetric technology / Non-radiometric technology
- Patient usage
- Adult and Pediatric
- Type of Culture Bottles Supported
- Aerobic, Anaerobic, Fungal
- Sample Capacity (Minimum)
- 200 bottles or more, 360 bottles or more Or higher
- Specification Name
- बड के िलए आव8यक अनुमत मूHय /Bid Requirement (Allowed Values)
- System should have the
- Yes
- System should be able to display growth kinetics on the screen and have the capacity to generate hard copy of each growth kinetics
- Yes
- System should have the facility to analyse delayed specimens with the routine bottles
- Yes
- System should be having continuous agitation and incubation facility to provide optimal growth of microorganisms
- Yes
- System should have in-built calibration check/facilitate calibration
- Yes
- Media bottles should have the capacity to neutralize antibiotics
- Yes
- Visual and audible alarms for positive cultures
- Yes
- Every cell (bottle position) have its own optics and detection device
- Yes
- System should have data management system and software to analyze and store the data
- Yes
- Barcode Reader
- Built-in / External
- Type of monitor
- Inbuilt / PC Based
- Type of monitor display
- LCD / LED
- Touch screen display
- Yes
- UPS provided for Power Backup facility
- Yes
- UPS Back up time in minutes
- 30 minutes or more, 60 minutes or more
- Availability of external printer
- No
- Connectivity
- LIS / HIS compatibility
- Compliance to Medical Device Rules (MDR) 2017 as amended till date
- Yes
- Specification Name
- बड के िलए आव8यक अनुमत मूHय /Bid Requirement (Allowed Values)
- Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date
- Yes
- Certification for manufacturing unit
- ISO:13485 (Latest)
- Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date
- Yes
- Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement
- Yes
- Conformity to electrical safety requirements
- IEC 60601 / 61010 latest amended or BIS equivalent
- Warranty in years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue)
- 1.0 (year)
- Comprehensive Maintenance ( Minimum Percentage )
- 3 %
- Comprehensive Maintenance ( Maximum Percentage )
- 25 %
- Warranty of required product
- 5 Year
- Comprehensive Maintenance Duration (Post Warranty)
- 5 Year
- Earnest Money Deposit (EMD)
- The EMD shall be submitted in the form of a Fixed Deposit Receipt (FDR) having a validity of one year. The EMD shall be submitted in physical form, strictly as per the terms and conditions prescribed on the GeM portal.
- Manufacturer’s Authorization
- In case the bidder is other than the original manufacturer, a valid authorization from the original manufacturer shall be submitted along with the bid.
- Demo/Trial
- The buyer reserves the right to call for trial/demonstration of the quoted equipment during the technical evaluation process. The bidder shall arrange the demonstration at its own cost, as permissible under the prevailing GeM Terms & Conditions. Failure to provide the demonstration, if requested, may result in rejection of the technical bid.
- Specification
- The firm should comply with the specification uploaded in the bid at Buyer added ATC failure of which the bid will be rejected.
- Software updates 2
- 1. The bidder should submit that they will provide software updates for the equipment free of cost. 2. The bidder will provide one printer with the machine along with warranty and CAMC as provided to the machine.