Digital Ultrasonic Flaw Detector
600043,NUCLEAR POWER CORPORATION OF INDIA LTDChennai QA Unit,DAE Nodal Centre, Plot-B, Main Gate-2, Dr. Ambedkar Road, Pallavaram, Chennai, Tamilnadu - 600043.Phone: 044 -22640003; 22644233; 22644011
- Delivery
- 14 days
Specifications & requirements · 7
- Specification Name
- बड के िलए आवKयक अनुमत मूRय /Bid Requirement (Allowed Values)
- Availability of ISO certification of manufacture
- Yes, No
- Test Mode 2.Display 3.Function Keys 4.Battery Life 5.Display dimensions 6.Transducer socket 7.Velocity 8.Resolution 9.Test Range 10.Pulser Width 11.Damping 12.Pulse repetition freq
- Pulse-echo, through transmission or dual Touch screen 22.8 cm (9 inch) TFT Touch 24 Hrs per charge (Lithium-iron-Phosphate) (W) 200 mm x (H) 110 mm LEMO-00 type connectors-Two 1000-8000 m/s 800 (RGB) x 480 25 mm to 6m in steel User Variable 120 ohms & 1000 ohm Selectable 50 Hz, 100 Hz, 1 kHz
- Instrument Bandwidth 14.Gain 15.Rectification 16.B-scan 17.Beam Plot 18.Scientific Calculator 19.System linearity 20.Measurement Gates 21.Measurement 22.Echo to Echo Measurement 23.Unit
- 1MHz – 18MHz 0-mini 95dB RF, Full wave YES To identify the discontinuity Location In-bult Horizontal: < 1> Two independent gates for measurement and monitoring Thickness, beam path, Depth, Amplitude, Distance With a single gate Selectable: mm or inches for range
- Internal memory 25.Interface 26.Screen Shot 27.Velocity 28.Display Delay 29.Evaluation Options 30.DAC 31.Video Recording 32.Power requirement 33.Operating temperature
- 32GB Connectivity to External PC/VGA YES 1000 m/s to 8000 m/s 0 to 2812 Microseconds TCG, DAC, DGS/AVG, AWD1.1(Auto Calculated D Value) Static & Dynamic 5 minutes continues & Max. 200Hrs. AC – mains, 200 – 240V (50-60 Hz) 0 to 50ºC
- Real time clock & Battery level indication shall be available. 35.Calibration Certificate and Test Certificate shall be supplied with the equipment. 36.Operation and maintenance manual shall be supplied with the equipment
- 34.Real time clock & Battery level indication shall be available. 35.Calibration Certificate and Test Certificate shall be supplied with the equipment. 36.Operation and maintenance manual shall be supplied with the equipment.
- Training: The Supplier shall provide comprehensive training on the operation and maintenance of the UT equipment to our Engineers. 38.Warranty: The supplier shall provide warranty for at least 12 months after commissioning of equipment
- 37.Training: The Supplier shall provide comprehensive training on the operation and maintenance of the UT equipment to our Engineers. 38.Warranty: The supplier shall provide warranty for at least 12 months after commissioning of equipment. 39.Mandatory accessories as per attached Technical specification.