SPECTROPHOTOMETER
389350,Gujarat Natural Farming Science University Near New Court, Jambudi Farm, Pavagadh Road, Halol-389350. Panchmahal, Gujarat (Bharat)
- Delivery
- 15 days
Specifications & requirements · 41
- Specification Name
- बड के िलए आवPयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Light Source
- Halogen Lamp, Deuterium Lamp
- Diffraction Grating (lines/mm)
- 1200
- Switching of Light Source
- Automatic
- Wavelength Setting And Scanning
- Automatic
- Type of Spectrophotometer
- UV - VISIBLE
- Type of Optical System
- Double Beam
- Monochromator Type
- Czerny-turner
- Specification Name
- बड के िलए आवPयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Detector Type
- Photodiode
- Spectral Bandwidth Type
- Fixed Bandwidth
- Source Wavelength - Minimum
- 190.0 (nanometer)
- Source Wavelength - Maximum
- 1100, 1500, 2000, 2500, 2700, 3000, 3200, 3300, 1650 Or higher (nanometer)
- Resolution
- 1.0 (nanometer)
- Wavelength Accuracy (+/-)
- 0.1 (nanometer)
- Bandwidth (nm)
- 1
- Focal length (mm)
- 0 to 49
- Photometric measurement modes
- Transmittance, Absorbance, Concentration (Wavelength, time), Reflectance, Energy
- Photometric Absorbance (Max) (Abs) (+/-)
- 4
- Photometric Absorbance Accuracy (Max) (+/-)
- 0.001 - 0.002 Or lower
- Photometric Transmittance (Max) (+/-)
- 100.0 (percent)
- Photometric Transmittance Accuracy (Max) (+/-)
- 0.0 - 0.5 Or lower (percent)
- Photometric Reflectance (Max) (+/-)
- 100.0 (percent)
- Display
- In-built display
- Display type
- LCD touchscreen
- Display Size
- 6.8 - 30.0 Or higher (inch)
- Test result printing
- Through external printer
- If inbuilt printer, Type of printer
- N/A
- Internal storage(GB)
- 32.0 - 80.0 Or higher
- Mximum Path Length of sample
- 100.0 (millimeter)
- Number of Cuvettes supplied-inclusive in the scope of supply
- 2
- Spares and Consumables inclusive in the scope of supply
- No
- Minimum Operating Temperature
- 10.0 (degree Celsius)
- Maximum Operating Temperature
- 35.0 (degree Celsius)
- Specification Name
- बड के िलए आवPयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Operating Humidity (RH) (%) at 40 degree C
- 80.0 (percent)
- Onsite Warranty
- 1.0 (year)
- Availability of Spares (at extra cost)
- 8.0 (year)
- Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements
- Yes
- Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use
- Yes
- Availability of Test Reports from Central Govt/NABL approved/ILAC acredited lab to prove conformity to the specification
- No