Ultrasound Machine (V2)
700091,Swasthya Sathi, GN-29, Sector V, Salt Lake
- Delivery
- 30 days
Specifications & requirements · 70
- Specification Name
- बड के िलए आवOयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Convex Probe
- 2.5 to 5 MHz (±2)
- Microconvex probe
- 4 to 8 MHz (±2)
- Linear array probe
- 5 to 10 MHz (±2)
- TVS Probe
- Not Provided
- Cardiac Probe
- Not Provided
- Hockey Stick Probe
- Not Provided
- Volume Convex Probe
- Not Provided
- Volume Linear Probe
- Not Provided
- Volume TVS Probe
- Not Provided
- Endorectal Probe
- Not Provided
- E-Breast Elastography technique
- No
- E-Thyroid Elastography technique
- No
- Liver Elastography technique
- No
- Gel Warmer
- No
- Dedicated 3-D/4-D
- No
- Contrast harmonic imaging with angio mode
- No
- Provision of Harmonic imaging in Linear Probes
- Yes
- Color Doppler imaging (CDI)
- Yes
- Color Doppler 3-D/4-D option
- No
- Specification Name
- बड के िलए आवOयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Power Doppler imaging (PDI)
- Yes
- Power Doppler 3-D/4-D option
- No
- Continuous wave
- Yes
- Triplex mode
- Yes
- Tissue Doppler imaging
- Yes
- Contrast-enhanced ultrasound (CEUS) option
- No
- Tissue synchronization, tissue velocity
- No
- Provision for higher sensitivity for low frequency doppler in all probes
- No
- Number of channels in color doppler system
- 50000 or more (if channels are being designated as digital channels)
- Live 3 D imaging in color doppler system
- No
- Colour Doppler system with all application packages ,Quad loop for serial studies with high frame
- No
- Provision of Tissue Colorization (B-Colour) for improved contrast resolution
- Yes
- Cine Loop Features
- Yes
- Frame grabber facility for post analysis
- Yes
- Number of transducer/probe ports
- 3
- System should be capable of generating real time live 3- D images
- No
- Contrast agent imaging
- No
- DICOM 3 Compliant
- Yes
- Automated B-mode (2-D) image
- Yes
- Specification Name
- बड के िलए आवOयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Automated CDI(Color Doppler Imaging) image optimization
- Yes
- Automated PW Doppler image optimization
- Yes
- Touch Screen
- No
- Monitor display size (inch)
- 21 or more Or higher
- Split screen
- No
- Single/dual monitors
- Single
- Max number frames
- >=1000
- Digital storage hard drive, GB
- >=200 or more
- DVD/CD Writer/USB
- No
- Type of processor (in PC)
- NA
- RAM capacity of PC in GB (hint: In case PC not provided put zero)
- NA
- Processor speed in GHz (PC)
- NA
- Frame Grabber incorporated(PC)
- No
- Connectivity between USG and PC
- No
- PC with all software inclusive interfaced with USG machine
- No
- Laser Color printer provided
- No
- Inkjet/Thermal printer provided
- No
- CD/DVD/USB Produced should be playable on any system
- No
- Glossy Colour Print Paper supplied
- NA
- Dicom Interfacability Dicom to PACS
- No
- Interfacability Dicom to RIS
- No
- Resettable over current breaker shall be fitted for protection
- No
- Power backup
- UPS
- Specification Name
- बड के िलए आवOयक अनुमत मूVय /Bid Requirement (Allowed Values)
- Back up time in minutes (minimum)
- 30.0 (minute)
- Suitable Servo Controlled Stabilizer/CVT should be provided
- No
- Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue)
- 3, 5 Or higher (year)
- Warranty
- 2 Years
- Cerificate
- The bidder/OEM should submit valid CDSCO Certificate/ Registration/License for both the manufacturer(s) and importer(s) as applicable (including probe)