Karl Fischer Titrator (V2)
176061,CSIR-Institute of Himalayan Bioresource Technology.Palampur-176061
- Delivery
- 45 days
Specifications & requirements · 10
- Specification Name
- बड के िलए आवMयक अनुमत मूTय /Bid Requirement (Allowed Values)
- Titration Method
- Volumetric Karl Fischer Titrator, Coulometric Karl Fischer Titration
- Measurement Range
- Volumetric 0.1 mg to 100 mg H?O, Coulometric 1 µg to 1000 mg H?O
- Accessories
- Dual Platinum Electrode, Titration Stand, Power Supply Adapter, Generator Electrode
- Warranty
- 1 (year)
- Instrument should have the capability of performing Coulometric Karl Fischer titration with electrochemical in-situ iodine generation
- 1. Instrument should have a polarizer measuring range of 0–2300 mV (or better) with resolution of 0.1 mV (or better). 2. Instrument should have a digital touchscreen display providing clear status indication, operating signals, command execution, and full instrument operation. 3. Instrument should have automatic sample detection, automatic initiation of titration, and fully automatic cell conditioning without requiring manual intervention.
- Instrument should have the facility for user-defined calculations and automatic termination (Stop option) of titration based on either predefined drift limits or time, ensuring accurate results under varying environmental conditions
- 1. The instrument should have the facility to connect oven autosampler for high throughput sampling 2. Instrument should have data management facilities via Ethernet, USB flash drive, and printer, with the capability to generate PDF reports containing date and time stamps and save them directly to a connected USB flash drive.
- Instrument should be supplied with a compatible printer and USB flash drive for printing/transferring measurement results and reports
- 1. Instrument should operate on an input voltage of 100–240 V AC ±10% at 50–60 Hz, with a 24 V DC output and maximum power consumption of 120 W.
- Instrument should have a moisture measurement range of 10 ?g to 200 mg of water with results repeatability up-to
- 2. Instrument should have an integrated internal magnetic stirrer for uniform sample mixing during titration.
- The complete system should have comprehensive warranty of minimum 1 year and should be supplied with NABL- accredited calibration certificate
- 3. The system should be supplied with all the required accessories and consumables (including coulometric cell, generator electrode, solvent pump, titration kit, molecular sieves, certified reference material for calibration and performance validation, hydranal water standard, KF reagent, printing paper compatible with printer, additional printer cartridge) for full functioning of the instrument and analysis.