High End Multipara Monitor (V3)
208002,LPS Institute of Cardiology, G.T. Road, Rawatpur, Near Gol Chauraha, Kanpur
- Delivery
- 30 days
Specifications & requirements · 22
- Specification Name
- बड के िलए आवOयक अनुमत मूWय /Bid Requirement (Allowed Values)
- Type of Monitor
- Pre- Configured, Modular
- Monitoring Parameters
- ECG, HR, Respiration Rate, SpO2, Pulse Rate, NIBP, Temperature, IBP
- Length of time of full disclosure of patient data in hr
- ? 24, ? 48, ? 72 Or higher
- Length of time of trending in hr
- ? 24, ? 48, ? 72 Or higher
- Number of leads of ECG
- 3, 5, 6, 10
- ECG Waveform number of leads display
- 7, 8, 12
- Respiration rate measurement range
- 0 - 100 bpm or more, 0 - 150 bpm or more
- Specification Name
- बड के िलए आवOयक अनुमत मूWय /Bid Requirement (Allowed Values)
- SPO2 measurement range
- 40 - 100%, 70 - 100%
- Systolic Blood pressure monitoring range
- 30 - 300 mmHg or more, 40 - 250 mmHg or more, 40 - 270 mmHg or more
- Diastolic Blood pressure monitoring range
- 10 - 200 mmHg or more, 30 - 220 mmHg or more
- Type of probe
- Skin, Rectal, Esophageal
- Number of channels for IBP
- 2, 3, 4, 8
- Sensor type
- Mainstream, Sidestream, Microstream, NA
- Display Size in inches
- 10, 12, 15, 17, 19 Or higher
- Traces of display
- >6, >10, >12, >14
- User Interface of display
- Rotary knob, Touchscreen, Touchscreen and Rotary knob Or higher
- Comprehensive Maintenance ( Minimum Percentage )
- 3 %
- Comprehensive Maintenance ( Maximum Percentage )
- 25 %
- Warranty of required product
- 5 Year
- Comprehensive Maintenance Duration (Post Warranty)
- 5 Year